Temperature Transmitter For Marine Type MBT 9110
The MBT 9110 temperature transmitter can be used for converting the Pt 100 / Pt 1000 ohm signal or the thermo couple mV signal to a 4 – 20 mA signal.
The MBT 9110 is a very durable and reliable temperature transmitter with all major marine approvals and can be used together with all our sensors based on Pt or Thermocouple technology.
MBT 9110 transmitters in the standard program are delivered in an enclosure, but are also used as an integrated part
of certain temperature sensors. The transmitter is available as standard or galvanic isolated with or without EEx approval.
Primary accuracy | Pt 100 < ± 0.3 °C Thermocouple type E, J, K, L, N, T, U < ± 1.0 °C Thermocouple type B, R, S < ± 2.0 °C |
Linearity error | < 0.1% FS |
Temp. coefficient | < ± 0.01% FS / °Kamb |
Response time | Programmable 1 – 60 sec. 1 sec. standard |
Cold solder point compensation (CJC) | < ± 1.0 °C |
Warm-up time | 5 min. |
Update time | 440 ms |
Effect of sensor cable resistance (3/4 wires) | < 0.002 ? / ? |
Kamb = Ambient temperature change
Electrical specifications
Supply voltage | 8 – 35 V d.c | |
EEx Galvanic supply voltage | 7.2 – 30 V d.c | |
Effect of supply voltage variation | < 0.005% FS / V d.c | |
Output | 4 – 20 mA current loop | |
Sensor fault indication | Namur NE43 upscale 23 mA | |
Input | Standard | Pt 100 (EN 60751) 2, 3 – wire -200 – 800 °C Pt 1000 (EN 60751) 2, 3 – wire |
Galvanic isolation | Pt 100 (EN 60751) 2, 3, 4 – wire -200 – 800 °C Pt 1000 (EN 60751) 2, 3 – wire Thermocouple (EN 60584) | |
Signal/noise ratio | Min. 60 dB | |
Max. lead cross-section | 1 × 1.5 mm² | |
Max. cable resistance | 5 ? / core |
Insulation voltage | 1500 V a.c |
EMC | EN 61326-1 |
Vibration/shock | IEC 60068-2-6 / IEC 60068-2-84 |
Vibration | 4 g / 2 –100 Hz |
Humidity | 0 – 98% RH, according to IEC 60068-1, IEC 60068-2-2 |
Ambient temperature | -40 – 70 °C |
In seperate enclosure | IP54 according to IEC 60529 |
Max. offset | 50% of max. input value (Pt 100: 400 °C) (Thermocouple, type R: 650 °C) |
Net weight | In heightened lid for B-head: 0.080 kg In Box enclosure: 0.360 kg |
Read more :
Rheonik Fuel Consumption For Marine Application, Rheonik Coriolis mass flow meters have been used for…
Coriolis Mass Flow Meters For Marine Fuel Consumption Application, Rheonik Coriolis mass flow meters have…
In commercial shipping, now more than ever, fuel consumption measurements are an important part of…
In commercial shipping, now more than ever, fuel consumption measurements are an important part of…
Mengukur konsumsi bahan bakar di sebuah mesin atau engine, disini kita melihat penjelasan bagaimana bahan…
Mengukur Konsumsi Bahan Bakar di Kapal Dengan Fluidwell Tipe F127 Mengukur konsumsi bahan bakar di…
This website uses cookies.
View Comments